Logo

Condensables_CI-APi-TOF_2_SMR

operational
🔬 chemical ionization-atmospheric pressure interface-time of flight mass spectrometer 📌 21.11157/p7ke6-z1243
▶ Description

Chemical ionization-atmospheric pressure interface-time of flight mass spectrometer operated at the observatory platform SMEAR II (Hyytiälä) to perform in-situ measurements of condensable vapours.


▶ 🔧 Inlet System from 9999-01-01
Operated in atmospheric simulation chamber ✗
Measurement height above ground 35.0 m
Distance from NOx inlet 5.0 m
Distance from VOC inlet 5.0 m
Distance from ozone inlet 5.0 m
Inlet tube material Stainless steel
Residence time 1.0 s
Inlet line heating ✗
Filter inlet material None
▶ 🧬 Compounds from 2016-01-01
#Name
1 sulfuric acid
2 methanesulfonic acid
3 iodic acid
4 highly oxidized molecules
▶ 🔬 Mass analyzer from 2016-01-01
Type Time-of-Flight
Time-of-Flight type H-TOF
Mass resolution 2500
Instrument manufacturer Aerodyne Research
Instrument model Aerodyne CI-API-TOF
Instrument serial number 941.16.0109
Reagent ions Nitrate
Switching time 0.0 min
Filter inlets for simultaneous measurements of gas and condensed phase HOM ✗
Pressure atmospheric
Sample flow 0.8 l/min
Total air flow 10.0 l/min
Sample flow treatment with ion filter ✗
▶ 📐 Calibration from 9999-01-01
Method Sulfuric acid calibrator
Frequency 2-3 times a year
▶ 📋 Measurement QA/QC
Instrument log book Electronic
Checklist 1.0 weeks
▶ ✅ Data QA/QC
Time series analysis of calibrations ✗
Time series analysis of compound ratios ✓
@VOC@ ✗
▶ 📤 Data Submission
Database Submission interval
ACTRIS –
▶ 👤 Operators
OperatorLevel
0000-0003-1874-3235 expert