operational
🔬 gas chromatograph, offline 🏭 PerkinElmer 📦 PerkinElmer Clarus 580 S/N 580S12020707
Description

Gas chromatograph operated at the observatory platform Schmücke to perform offline in-situ measurements of non-methane hydrocarbons (NMHCs)


🔧 Inlet System from 9999-01-01
Operated in atmospheric simulation chamber
Measurement height above ground 18.0 m
Measurement height above building 2.0 m
Distance from NOx inlet 0.55 m
Distance from ozone inlet 0.65 m
Inlet tube material PFA/stainless steel
Inlet line heating
Inlet line temperature 25.0 °C
Filter inlet material None
💨 Sample from 9999-01-01
Canister sampling Pumping
Pump
Kind of pump Autocan
Pump manufacturer Breitfuss
Air flow 500.0 mL/min
Time 5.0 minutes
Air sampling volume 1.8 litre
Sampling interval 2.0 days
Canister volume 1.8 litre
Canister pressure pre-sampling 2.0 bar
Canister pressure post-sampling 2.0 bar
Storage time 2.0 weeks
🧬 Compounds from 9999-01-01
#Name
1 (E)-2-butene
2 (Z)-2-butene
3 1,3-butadiene
4 1-butene
5 2-methylbutane
6 2-methylpentane
7 2-methyl-propane
8 benzene
9 ethane
10 ethene
11 ethylbenzene
12 ethyne
13 isoprene
14 m-xylene
15 n-butane
16 n-heptane
17 n-hexane
18 n-pentane
19 o-xylene
20 p-xylene
21 propane
22 propene
23 toluene
24 2-methyl-propene
🧪 Air Treatment during canister sampling from 9999-01-01
Water removal None
CO₂ removal None
Oxidant removal None
Particle removal None
💉 Introduction device from 9999-01-01
Technique Concentrating on adsorbents
Concentrating on adsorbents
Adsorption material Air Monitoring Trap
Adsorption temperature -30.0 °C
Inlet mode Thermodesorption
Thermodesorption mode not specified
Air sampling system manufacturer PerkinElmer
Sampling system serial number TD300l1203024
Sampling flow 15.0 mL/min
Air sampling duration 33.3 min
Injection volume 1.8 L
Injection split Splitless
Injection pressure 1.5 bar
Number of replicates 1.0
⚗️ GC System from 9999-01-01
Instrument manufacturer PerkinElmer
Instrument model PerkinElmer Clarus 580
Instrument serial number 580S12020707
First column
Length 50.0 m
Material CP-AI2O3 NA2SO4
Film thickness 5.0 µm
Inner diameter 320.0 µm
Second column
Length 50.0 m
Material BP1
Film thickness 1.0 µm
Inner diameter 220.0 µm
📡 Detection from 9999-01-01
Detection method FID
Detection system
Manufacturer MICROTEC
Model MICROTEC REV C
Serial number N610-0374
📐 Calibration from 9999-01-01
Zero gas generation technique Catalyst
Zero gas generation with catalyst
Catalyst type zero air generator NBS
Catalyst temperature 450.0 °C
Calibration frequency 1.0 months
Zero gas / Blank frequency 2.0 weeks
Linearity check frequency 1.0 months

Certificates (1)

Name: NPL_NMHCs_2023_WAL Supplier: NPL Cylinder: D982550 Valid: 2023-02-03 → 2028-02-02
📋 Measurement QA/QC
Instrument log book Electronic
Checklist 1.0 days
✅ Data QA/QC
Time series analysis of calibrations
Time series analysis of compound ratios
@VOC@
📤 Data Submission
Database Submission interval
ACTRIS 1.0 years
EMEP 1.0 years