NMHCs_GC_offline_CN18393134_KOS

operational
🔬 gas chromatograph, offline 🏭 Agilent 📦 Agilent 7890B S/N CN18393134
Description

Gas chromatograph operated at the observatory platform National Atmospheric Observatory Kosetice to perform offline in-situ measurements of non-methane hydrocarbons (NMHCs).


🔧 Inlet System from 2025-11-01
Operated in atmospheric simulation chamber
Measurement height above ground 1.4 m
Distance from NOx inlet 50.0 m
Distance from ozone inlet 50.0 m
Inlet tube material PFA
Residence time 1.0 s
Inlet line heating
Filter inlet material PTFE (Teflon)
Lower mesh size 10.0 µm
Upper mesh size 10.0 µm
💨 Sample from 2012-01-01
Canister sampling Pumping
Pump
Kind of pump Membrane
Pump manufacturer KNF
Air flow 4750.0 mL/min
Time 2.0 minutes
Canister type Electropolished stainless steel canisters
Air sampling volume 0.4 litre
Sampling interval 72.0 hours
Canister volume 1.5 litre
Canister pressure pre-sampling 0.001 bar
Canister pressure post-sampling 2.0 bar
Number of replicates 1.0
Storage time 30.0 days
🧬 Compounds from 2022-04-01
#Name
1 ethane
2 ethene
3 propane
4 propene
5 2-methyl-propane
6 n-butane
7 ethyne
8 (E)-2-butene
9 (Z)-2-butene
10 1-butene
11 2-methylbutane
12 n-pentane
13 1,3-butadiene
14 (2E)-2-pentene
15 1-pentene
16 2-methylpentane
17 n-hexane
18 isoprene
19 benzene
20 2,2,4-trimethylpentane
21 n-heptane
22 toluene
23 n-octane
24 ethylbenzene
25 m/p-xylenes
26 o-xylene
27 1,3,5-trimethylbenzene
28 1,2,4-trimethylbenzene
29 1,2,3-trimethylbenzene
🧪 Air Treatment during canister sampling from 2012-01-01
Water removal None
CO₂ removal None
Oxidant removal None
Particle removal PTFE (Teflon)
💉 Introduction device from 2022-04-01
Technique Concentrating on adsorbents
Concentrating on adsorbents
Adsorption material Markes Ozone precursor trap
Adsorption temperature -25.0 °C
Desorption temperature 300.0 °C
Inlet mode Thermodesorption
Thermodesorption mode Direct desorption
Air sampling system manufacturer Markes
Air sampling system model Markes UNITY-xr
Sampling system serial number GB00U33126-18/11
Sampling flow 25.0 mL/min
Air sampling duration 16.0 min
Injection volume 0.4 L
Injection interval 60.0 min
Injection split Splitless
Injection pressure 2.0 bar
Number of replicates 1.0
⚗️ GC System from 2022-04-01
Instrument manufacturer Agilent
Instrument model Agilent 7890B
Instrument serial number CN18393134
First column
Length 50.0 m
Material Agilent J&W HP-AL/S
Film thickness 8.0 µm
Inner diameter 320.0 µm
Second column
Length 60.0 m
Material Agilent J&W HP-1 
Film thickness 1.0 µm
Inner diameter 250.0 µm
📡 Detection from 2022-04-01
Detection method FID
Detection system
Manufacturer Agilent
Model Agilent 7890B
Serial number CN18393134
📐 Calibration from 2022-04-01
Zero gas generation technique cylinder gas N2 (6.0)
Calibration frequency 3.0 months
Number of repetitions 1.0
Zero gas / Blank frequency 1.0 days
Blank substraction Not applied
Linearity check frequency 3.0 months

Certificates (2)

Name: NPL_NMHC_2021_KOS Supplier: NPL Cylinder: D933541 Valid: 2021-11-23 → 2026-11-22 Applied: 2022-04-01 → ?
Name: NPL_NMHC_2022_KOS Supplier: NPL Cylinder: D982552 Valid: 2022-06-22 → 2027-06-24 Applied: 2022-04-01 → ?
📋 Measurement QA/QC
Instrument log book Hard copy
Checklist 1.0 days
✅ Data QA/QC
Time series analysis of calibrations
Time series analysis of compound ratios
@VOC@
📤 Data Submission
Database Submission interval
ACTRIS 1.0 years
👤 Operators
OperatorLevel
0009-0001-7904-5548 scientist
0009-0001-8578-9797 expert
Jan Čech technician
0009-0000-7997-9642 technician