Logo

NOx_CAPS_xx_HPB

operational
🔬 chemiluminescence detector 🏭 Aerodyne Research

📝 Notes

* Please check the remarks. Do all of them really belong to the CAPS instrument? * Please select an instrument model from the list. If not available send and email to k.seemeyer@fz-juelich.de * Inlet line temperature: T_ambient + 4°C * Be aware that "Duration of stay in converter or bypass line" 0.6 s is rounded up to 1 s by document. * NO2 CAPS is attached collocated with CLD1 on inlet line and calibrated contemporaneously. * The second Database is the internal MIRAKEL PRJ MOHP Database (Oracle, SQL). * pressure: Vaisala, PTB330 * humidity: E+E Elektronik GmbH, EE33 * for the ozone instrument and metereology instrument the same modules as for https://cigas-box.fz-juelich.de/login/objects/75 were added, if this is not correct add new ozone and metereology instruments and add those.

🔧 Inlet System
Residence time 7.0 s
Inlet tube material PFA
Filter inlet material PTFE (Teflon)
Inlet filter mesh size 5.0 µm
Distance from voc inlet 6.0 m
Inlet line heating bool
Distance from ozone inlet 0.0 m
Humidity temperature control Temperature controlled
Measurement height above ground 16.0 m
Distance from condensables inlet 11.0 m
📐 Calibration
No2
Type GPT
Frequenzy 6.0 hours
No2 model [Referenz-ID: 938]
Automation Automatic
Certificates
Certificate
Certificate [Referenz-ID: 451]
Certificate [Referenz-ID: 453]
Certificate [Referenz-ID: 450]
Certificate [Referenz-ID: 452]
No2 manufacturer [Referenz-ID: 937]
No2 serial number
Interval 95%
No check
No2 check
Uncertainty 0.04 µmol/mol
Concentration 10.0 µmol/mol
📤 Data Submission
Databases
Database Submission
ACTRIS 1.0 years
✅ Data QA/QC
Bool voc
Bool other
Bool ratios
Bool calibrations
📋 Measurement QA/QC
Checklist 1.0 days
Instrument log book Electronic