NOx_CLD_1_HPB

Ecophysics_CLD770ALppt_CLD1

operational
🔬 chemiluminescence detector 🏭 Ecophysics 📦 Ecophysics CLD700ALppt S/N 00903392 (Inv# VDE) 📌 http://hdl.handle.net/21.11157/dx4dm-3e964
Description

Chemiluminescence detector operated at the observatory platform Meteorological Observatory Hohenpeissenberg (DWD) to perform in-situ measurements of nitrogen oxides (NOx).


🔧 Inlet System
Operated in atmospheric simulation chamber
Measurement height above ground 16.0 m
Measurement height above building 1.86 m
Distance from VOC inlet 6.0 m
Distance from condensables inlet 11.0 m
Distance from ozone inlet 0.0 m
Inlet tube material PFA
Residence time 5.0 s
Humidity/Temperature control Temperature controlled
Inlet line heating
Filter inlet material PTFE (Teflon)
Inlet filter mesh size 5.0 µm
⇄ Converter
Type Photolytic
Converter manufacturer Air Quality Design
Converter model Air Quality Design BLC/Custom
Converter serial number 437418 (DWD inventory number)
Duration of stay in converter or bypass line 1.0 s
Duration of stay in converter 1.0 s
Temperature 300.0 K
📐 Calibration
Gas standard concentration 10.0 µmol/mol
Gas standard uncertainty 0.04 µmol/mol
Gas standard confidence interval 95%
Zero air Zero Air generator
Zero air manufacturer Ecophysics
Zero air model Ecophysics PAG003

NO Calibration

Type Cylinder/Dilution
Automation Automatic
Frequency 6.0 hours
NO calibrator manufacturer MLU
NO calibrator model MLU AirQrate

NO Certificates (4)

Name: NPL_NOx_2008_HPB Supplier: NPL Cylinder: 195524 SG Valid: 2008-10-14 → 2009-10-13
Name: NPL_NOx_2012_HPB Supplier: NPL Cylinder: P219527 Valid: 2012-01-26 → 2014-01-25
Name: NPL_NOx_2019_HPB Supplier: NPL Cylinder: 2795 Valid: 2019-08-30 → 2021-08-29 Applied: 2018-08-23 → 2018-12-19
Name: NPL_NOx_2024_HPB Supplier: NPL Cylinder: D180691 Valid: 2024-11-12 → 2026-11-11 Applied: 2018-12-19 → 2020-01-15

NO₂ Calibration

Type GPT
Automation Automatic
Frequency 6.0 hours
NO₂ calibrator manufacturer MLU
NO₂ calibrator model MLU AirQrate

NO₂ Certificates (4)

Name: NPL_NOx_2008_HPB Supplier: NPL Cylinder: 195524 SG Valid: 2008-10-14 → 2009-10-13
Name: NPL_NOx_2012_HPB Supplier: NPL Cylinder: P219527 Valid: 2012-01-26 → 2014-01-25
Name: NPL_NOx_2019_HPB Supplier: NPL Cylinder: 2795 Valid: 2019-08-30 → 2021-08-29
Name: NPL_NOx_2024_HPB Supplier: NPL Cylinder: D180691 Valid: 2024-11-12 → 2026-11-11
📏 Auxiliary Measurements
Role Type Manufacturer Model Serial number
Ozone instrument ultra-violet absorption detector Thermo Fisher Scientific Thermo 49C
Meteorology instrument - Humidity E+E Elektronik E+E Elektronik EE33
Meteorology instrument - Pressure Vaisala Vaisala PTB330
📋 Measurement QA/QC
Instrument log book Electronic
Checklist 1.0 days
✅ Data QA/QC
Time series analysis of calibrations
Time series analysis of compound ratios
@VOC@
📤 Data Submission
Database Submission interval
ACTRIS 1.0 years
👤 Operators
OperatorLevel
0000-0002-6445-9510 scientist