NOx_CLD_xx_CMN
operational
🔬 chemiluminescence detector
🏭 Teledyne
📦 Teledyne T200UP
▶ 📝 Notes
* converter: Integrated in CLD * The Teledyne T200UP was installed at Mt. Cimone on 01/08/2023. Prior to it, there was a Thermo Scientific 42iTL equipped with BLC, operating since January 2015. A future upgrade plan will be to measure the temperature, pressure, and relative humidity within the sampling system, and the active control of the flow.
▶ 🔧 Inlet System
| Residence time | 2.5 s |
| Inlet tube material | Other |
| Filter inlet material | PTFE (Teflon) |
| Inlet filter mesh size | 5.0 µm |
| Distance from voc inlet | 0.0 m |
| Inlet line heating bool | ✗ |
| Distance from ozone inlet | 0.0 m |
| Inlet tube material if other | Teflon and Pyrex |
| Measurement height above ground | 8.0 m |
▶ 📐 Calibration
| No | |||||
| Type | GPT | ||||
| No model | [Referenz-ID: 931] | ||||
| Frequenzy | 7.0 days | ||||
| Automation | Automatic | ||||
Certificates
|
|||||
| No manufacturer | [Referenz-ID: 603] | ||||
| No serial number | |||||
| No2 | |||||
| Type | GPT | ||||
| Frequenzy | 7.0 days | ||||
| No2 model | [Referenz-ID: 931] | ||||
| Automation | Automatic | ||||
Certificates
|
|||||
| No2 manufacturer | [Referenz-ID: 603] | ||||
| No2 serial number | |||||
| Interval | 95% | ||||
| No check | ✓ | ||||
| Zero air | Zero Air generator | ||||
| No2 check | ✓ | ||||
| Uncertainty | 0.05 µmol/mol | ||||
| Concentration | 5.0 µmol/mol | ||||
| Zero air model | [Referenz-ID: 940] | ||||
| Zero air manufacturer | [Referenz-ID: 603] | ||||
▶ 📤 Data Submission
Databases
|
|||||||
▶ ✅ Data QA/QC
| Bool voc | ✗ |
| Bool other | ✓ |
| Bool ratios | ✗ |
| Select if other | Automatic procedures for both operations |
| Bool calibrations | ✗ |
▶ 📋 Measurement QA/QC
| Checklist | 1.0 days |
| Instrument log book | Electronic |