Logo

NOx_CLD_photolytic_COPDD

operational
🔬 chemiluminescence detector 🏭 Teledyne 📦 Teledyne T200UP 📌 21.11157/z1kh4-nax89
▶ Description

Chemiluminescence detector operated at the observatory platform Site d’observation atmosphériques Puy de Dôme/ Opme/Cézeaux (COPDD) to perform in-situ measurements of nitrogen oxides (NOx).


▶ 🔧 Inlet System
Operated in atmospheric simulation chamber ✗
Measurement height above ground 12.0 m
Distance from VOC inlet 0.0 m
Distance from ozone inlet 0.0 m
Inlet tube material PTFE (Teflon)
Residence time 2.0 s
Humidity/Temperature control Temperature conditioning at 20 °C
Inlet line heating ✗
Filter inlet material PTFE (Teflon)
Inlet filter mesh size 5.0 µm
▶ ⇄ Converter
Type Photolytic
Converter manufacturer Teledyne
Converter serial number 198
Duration of stay in converter or bypass line 1.0 s
Duration of stay in converter 1.0 s
Temperature 293.15 K
▶ 📐 Calibration
Gas standard concentration 5013.0 µmol/mol
Gas standard uncertainty 100.0 µmol/mol
Gas standard confidence interval 2%
Zero air Zero Air generator

NO Calibration

Type Cylinder/Dilution
Automation Manual
Frequency 1.0 months
NO calibrator manufacturer Thermo Fisher Scientific
NO calibrator model Thermo Fischer Scientific 146-i

NO Certificates (1)

Name: Air_liquide_NOx_2025_COPDD Supplier: Other Cylinder: H5FW09G Valid: 2025-11-26 → 2026-11-26

NO₂ Calibration

Type GPT
Automation Manual
Frequency 1.0 months
NO₂ calibrator manufacturer Thermo Fisher Scientific
NO₂ calibrator model Thermo Fischer Scientific 146-i

NO₂ Certificates (1)

Name: Air_liquide_NOx_2025_COPDD Supplier: Other Cylinder: H5FW09G Valid: 2025-11-26 → 2026-11-26
▶ 📏 Auxiliary Measurements
Role Type Manufacturer Model Serial number
Ozone instrument ultra-violet absorption detector Thermo Fisher Scientific Thermo 49i 1161650015
Meteorology instrument – Vaisala Vaisala custom built –
▶ 📋 Measurement QA/QC
Instrument log book Hard copy
Checklist 1.0 weeks
▶ ✅ Data QA/QC
Time series analysis of calibrations ✗
Time series analysis of compound ratios ✗
@VOC@ ✗
Other QA/QC method Outlier detection
▶ 📤 Data Submission
Database Submission interval
ACTRIS 1.0 years