NO2_CAPS_146_BRM

operational
🔬 cavity attenuated phase shift spectrometer 🏭 Teledyne 📦 Teledyne T500U S/N 146

📝 Notes

* Zero air manufacturer: self made, air compressor: KNF and self construction of charcoal and Purafil cartridge, air compressor: N145.2ANE or PM15486-145.2 * This sheet only concerns NO2! * Zero air generator: air compressor with self made charcoal and Purafil cartridge, and CO to CO2 converter * NO2 permeation tube is used as a transfer to link NO2-Instrument (CAPS) to CLD (42i-TL) that is calibrated with NO reference gas. * Gas Standard Concentration and Gas Standard Uncertainty are taken from the scale certificate

🔧 Inlet System
Residence time 2.6 s
Inlet tube material PTFE (Teflon)
Filter inlet material Other
Inlet filter if other instrument internal
Distance from voc inlet 0.0 m
Inlet line heating bool
Distance from ozone inlet 0.0 m
Humidity temperature control None
Measurement height above ground 5.0 m
Distance from condensables inlet 0.0 m
📐 Calibration
No2
Type Cylinder/Dilution
Frequenzy 4.0 weeks
No2 model [Referenz-ID: 921]
Automation Manual
Certificates
Certificate
Certificate [Referenz-ID: 457]
No2 manufacturer [Referenz-ID: 574]
No2 serial number 904635
Interval not available
No check
Zero air Zero Air generator
No2 check
Uncertainty 0.1 µmol/mol
Concentration 20.07 µmol/mol
Zero air manufacturer [Referenz-ID: 886]
📤 Data Submission
Databases
Database Submission
ACTRIS 1.0 years
✅ Data QA/QC
Bool voc
Bool other
Bool ratios
Bool calibrations
📋 Measurement QA/QC
Checklist 2.0 weeks
Instrument log book Electronic