Logo

NOx_CAPS_N58_SMU

operational
🔬 cavity attenuated phase shift spectrometer 🏭 Envea 📦 Envea AC32M S/N N58

📝 Notes

* KS: Certificate: Supplier name and pdf not provided * Comment on Inlet line heating (row 26): ambient air inlet line is heated to 5 K above the ambient temperature. The laboratory is air-conditioned to 25 °C.

🔧 Inlet System
Line heating 25.0 degC
Residence time 2.0 s
Inlet tube material Other
Filter inlet material PTFE (Teflon)
Inlet filter mesh size 5.0 µm
Distance from voc inlet 0.55 m
Inlet line heating bool
Distance from ozone inlet 0.1 m
Humidity temperature control Temperature controlled
Inlet tube material if other PFA/stainless steel
Measurement height above ground 18.0 m
📐 Calibration
No2
Type Permeation
Frequenzy 25.0 hours
No2 model [Referenz-ID: 915]
Automation Automatic
Certificates
No2 manufacturer [Referenz-ID: 914]
No2 serial number 15104-448
No check
No2 check
📤 Data Submission
Databases
Database Submission
EMEP 1.0 years
ACTRIS 1.0 years
✅ Data QA/QC
Bool voc
Bool other
Bool ratios
Bool calibrations
📋 Measurement QA/QC
Checklist 1.0 days
Instrument log book Electronic