NOx_CLD_56_SIRTA

operational
🔬 chemiluminescence detector 🏭 Teledyne 📦 Teledyne T200UP S/N 56

📝 Notes

* Meteorology instrument model: Probe PT100 / Barometer PTB330 / Radiometer UV-S-AE-T

🔧 Inlet System
Line heating 21.0 degC
Residence time 3.33 s
Inlet tube material PTFE (Teflon)
Filter inlet material Other
Inlet filter if other 47 mm filter membrane- 1-2 micron
Distance from voc inlet 0.65 m
Inlet line heating bool
Distance from ozone inlet 0.0 m
Humidity temperature control Nafion dryer
Measurement height above ground 10.0 m
Measurement height above building 1.5 m
📐 Calibration
No
Type Cylinder/Dilution
No model [Referenz-ID: 919]
Frequenzy 1.0 months
Automation Manual
Certificates
Certificate
Certificate [Referenz-ID: 459]
No manufacturer [Referenz-ID: 601]
No serial number 86
No2
Type GPT
Frequenzy 3.0 months
No2 model [Referenz-ID: 919]
Automation Manual
Certificates
Certificate
Certificate [Referenz-ID: 459]
No2 manufacturer [Referenz-ID: 601]
No2 serial number 86
Interval 95%
No check
Zero air Zero Air generator
No2 check
Uncertainty 0.06 µmol/mol
Concentration 10.0 µmol/mol
Zero air model [Referenz-ID: 972]
Zero air manufacturer [Referenz-ID: 601]
📤 Data Submission
Databases
Database Submission
ACTRIS 1.0 years
✅ Data QA/QC
Bool voc
Bool other
Bool ratios
Bool calibrations
📋 Measurement QA/QC
Checklist 1.0 weeks
Instrument log book Electronic