NOx_CLD_56_SIRTA

operational
🔬 chemiluminescence detector 🏭 Teledyne 📦 Teledyne T200UP S/N 56
Description

Chemiluminescence detector operated at the observatory platform Site Instrumental de Recherche par Télédétection Atmosphérique (SIRTA) to perform in-situ measurements of nitrogen oxides (NOx).


🔧 Inlet System
Operated in atmospheric simulation chamber
Measurement height above ground 10.0 m
Measurement height above building 1.5 m
Distance from VOC inlet 0.65 m
Distance from ozone inlet 0.0 m
Inlet tube material PTFE (Teflon)
Residence time 3.33 s
Humidity/Temperature control Nafion dryer
Inlet line heating
Inlet line temperature 21.0 °C
Filter material 47 mm filter membrane- 1-2 micron
Inlet filter mesh size 2.0 µm
⇄ Converter
Type Photolytic
Converter manufacturer Air Quality Design
Converter model Air Quality Design Standalone BLC AQD low pressure
Converter serial number C-BLC-0186
Duration of stay in converter or bypass line 0.005 s
Duration of stay in converter 1.0 s
Temperature 294.0 K
📐 Calibration
Gas standard concentration 10.0 µmol/mol
Gas standard uncertainty 0.06 µmol/mol
Gas standard confidence interval 95%
Zero air Zero Air generator
Zero air manufacturer Teledyne
Zero air model Teledyne 701H

NO Calibration

Type Cylinder/Dilution
Automation Manual
Frequency 1.0 months
NO calibrator manufacturer Teledyne
NO calibrator model Teledyne T700U
NO calibrator serial number 86

NO Certificates (1)

Name: NPL_NOx_2013_SIRTA Supplier: NPL Cylinder: 1635 Valid: 2014-01-01 → 2025-12-01 Applied: 2014-01-01 → 2025-12-01

NO₂ Calibration

Type GPT
Automation Manual
Frequency 3.0 months
NO₂ calibrator manufacturer Teledyne
NO₂ calibrator model Teledyne T700U
NO₂ calibrator serial number 86

NO₂ Certificates (1)

Name: NPL_NOx_2013_SIRTA Supplier: NPL Cylinder: 1635 Valid: 2014-01-01 → 2025-12-01 Applied: 2014-01-01 → 2025-12-01
📏 Auxiliary Measurements
Role Type Manufacturer Model Serial number
Ozone instrument ultra-violet absorption detector Teledyne Teledyne T400 6168
Meteorology instrument - Temperature GUILCOR GUILCOR custom-made
Meteorology instrument - Pressure Vaisala Vaisala custom built
Meteorology instrument - UVA Kipp & Zonen
📋 Measurement QA/QC
Instrument log book Electronic
Checklist 1.0 weeks
✅ Data QA/QC
Time series analysis of calibrations
Time series analysis of compound ratios
@VOC@
📤 Data Submission
Database Submission interval
ACTRIS 1.0 years
👤 Operators