Logo

NOx_CLD_xx_SMR

operational
🔬 chemiluminescence detector 🏭 Thermo Fisher Scientific 📦 Thermo 42i-TL

📝 Description
🔧 Inlet System
Measurement height above ground 34.0 m
Distance from VOC inlet 0.0 m
Distance from condensables inlet 5.0 m
Distance from ozone inlet 0.0 m
Inlet tube material PTFE (Teflon)
Residence time 36.0 s
Humidity/Temperature control Other
Humidity/Temperature control Outdoors inlet + line heated at 5 C above air temperature; Indoors passive heating from atmospheric to lab temperature
Inlet line heating
Line heating 5.0 °C
Filter inlet material PTFE (Teflon)
Inlet filter mesh size 5.0 µm
📐 Calibration
NO calibration
NO Calibration
Type Other
Type if other NO from cylinder; NO2 generated via GPT of a 50 ppm NO standard in excess of O3
Automation Manual
Frequenzy 180.0 days
NO calibrator manufacturer Thermo Fisher Scientific
NO calibrator model Thermo Fischer Scientific 146-i
Certificates
Certificate
 → certificate → certificate
Certificate Linde_NOx_2019_SMR
NO2 calibration
NO2 Calibration
Type GPT
Automation Manual
Frequenzy 180.0 days
NO2 calibrator manufacturer Thermo Fisher Scientific
NO2 calibrator model Thermo Fischer Scientific 146-i
Certificates
Certificate
 → certificate → certificate
Certificate Air Products_NOx_2021_SMR
Concentration 1.04 µmol/mol
Uncertainty 0.01 µmol/mol
Interval 95%
Zero air Zero Air generator
Zero air manufacturer undefined
📤 Data Submission
Databases
Database Submission
ACTRIS 1.0 years
✅ Data QA/QC
Time series analysis of calibrations
Time series analysis of compound ratios
@VOC@
Bool other
📋 Measurement QA/QC
Instrument log book Electronic
Checklist 1.0 weeks