Logo

NOx_CLD_xx_SMR

operational
🔬 chemiluminescence detector 🏭 Thermo Fisher Scientific 📦 Thermo 42i-TL

📝 Notes

* zero air manufacturer/model: self made: compressed and dried air fed at a pressure of 2 bar through cartridges of Purafil + silicagel + activated + Sofnocat chemical * "scrubber" missing * checklist: including monitoring of status/diagnostic signals * sample inlets at 8 heights in a 125 m high guyed mast; data from 33.6 m height submitted, sampling/measuring time 1 min @ 6 min interval; sample line length 150 m, lag time 35 s; sample line 16/14 mm PTFE; Inlet type: Hat or hood Inlet description: Inlet protected against rain with indented and downward inclined apertures in a machined piece of a PTFE rod; 157 m long sample line of 16/14 mm diam. PTFE tube; flow rate 45 l/min; 5 um PTFE filter at analyser inlet Ozone correction: Submitted data corrected for reaction with O3 in the inlet * Water vapor correction: Submitted data corrected for water vapor quenching in CLD

🔧 Inlet System
Line heating 5.0 degC
Residence time 36.0 s
Humidity if other Outdoors inlet + line heated at 5 C above air temperature; Indoors passive heating from atmospheric to lab temperature
Inlet tube material PTFE (Teflon)
Filter inlet material PTFE (Teflon)
Inlet filter mesh size 5.0 µm
Distance from voc inlet 0.0 m
Inlet line heating bool
Distance from ozone inlet 0.0 m
Humidity temperature control Other
Measurement height above ground 34.0 m
Distance from condensables inlet 5.0 m
📐 Calibration
No
Type Other
No model [Referenz-ID: 931]
Frequenzy 180.0 days
Automation Manual
Certificates
Certificate
Certificate [Referenz-ID: 462]
Type if other NO from cylinder; NO2 generated via GPT of a 50 ppm NO standard in excess of O3
No manufacturer [Referenz-ID: 603]
No2
Type GPT
Frequenzy 180.0 days
No2 model [Referenz-ID: 931]
Automation Manual
Certificates
Certificate
Certificate [Referenz-ID: 463]
No2 manufacturer [Referenz-ID: 603]
Interval 95%
No check
Zero air Zero Air generator
No2 check
Uncertainty 0.01 µmol/mol
Concentration 1.04 µmol/mol
Zero air manufacturer [Referenz-ID: 886]
📤 Data Submission
Databases
Database Submission
ACTRIS 1.0 years
✅ Data QA/QC
Bool voc
Bool other
Bool ratios
Bool calibrations
📋 Measurement QA/QC
Checklist 1.0 weeks
Instrument log book Electronic