NOx_CLD_xx_PAS

operational
🔬 chemiluminescence detector 🏭 Thermo Fisher Scientific 📦 Thermo 42i-TL

📝 Notes

* zero air manufacturer: purified room air * Gas standard uncertainty: MU depends on dilution concentration (1.0 - 2.0 nmol/mol) * Checklist: Daily checks remotely during weekdays+station visit at least once a week * for the ozone instrument and metereology instrument the same modules as for https://cigas-box.fz-juelich.de/login/objects/60 were added, if this is not correct add new ozone and metereology instruments and add those.

🔧 Inlet System
Residence time 2.0 s
Inlet tube material Other
Filter inlet material None
Distance from voc inlet 0.0 m
Inlet line heating bool
Distance from ozone inlet 0.0 m
Humidity temperature control None
Inlet tube material if other Stainless steel/Silco steel
Measurement height above ground 7.0 m
📐 Calibration
No
Type Cylinder/Dilution
No model [Referenz-ID: 930]
Frequenzy 120.0 days
Automation Manual
Certificates
Certificate
Certificate [Referenz-ID: 471]
No manufacturer [Referenz-ID: 929]
No2
Type GPT
Frequenzy 12.0 months
No2 model [Referenz-ID: 930]
Automation Manual
Certificates
Certificate
Certificate [Referenz-ID: 471]
No2 manufacturer [Referenz-ID: 929]
No check
Zero air Zero Air generator
No2 check
Concentration 15.06 µmol/mol
📤 Data Submission
Databases
Database Submission
EMEP 1.0 years
✅ Data QA/QC
Bool voc
Bool other
Bool ratios
Bool calibrations
📋 Measurement QA/QC
Checklist 1.0 days
Instrument log book Electronic