NMHCs_GC_online_retired_CMN

retired
🔬 gas chromatograph, online

📝 Notes

**SCRIPPS certificate not provided** * Calibration frequency: SCRIPPS standard analyzed every two real sample; NPL std analyzed montly * Linearity check frequency: yearly or when any analytical condition has changed

💨 Sample from 9999-01-01
Pump
Pump model [Referenz-ID: 900]
Kind of pump Membrane
Pump manufacturer [Referenz-ID: 576]
Air flow 40.0 mL/min
Sample air flow select Vacuum sampling
🔧 Inlet System from 9999-01-01
Line heating 20.0 degC
Residence time 6.0 s
Lower mesh size 7.0 µm
Inlet tube material Other
Filter inlet material Stainless steel
Distance from nox inlet 0.0 m
Inlet line heating bool
Distance from ozone inlet 0.0 m
Inlet tube material if other composed inlet: 1 m of Pyrex (main inlet for all the instruments) then SS tube
Measurement height above ground 2.0 m
Measurement height above building 7.0 m
⚗️ GC System from 9999-01-01
First column
Length 60.0 m
Material Agilent GasPro
Inner diameter 320.0 µm
Instrument model [Referenz-ID: 902]
Pre column select box
Instrument manufacturer [Referenz-ID: 547]
Second column select box
📡 Detection from 9999-01-01
Select Mass spectrometry
Mass spectrometry
Type Quadrupole
Tune freq Monthly
Mass spectrometry model [Referenz-ID: 903]
Mass spectrometry manufacturer [Referenz-ID: 547]
🌬️ Air Sampling from 9999-01-01
Sampling volume 1.0 L
Sampling duration 25.0 min
Sampling system model [Referenz-ID: 901]
Air sampling technique Concentrating on adsorbents
Concentrating on adsorbents
Adsorption material MARKES U-T503F-2S
Adsorption temperature -35.0 degC
Desorption temperature 310.0 degC
Sampling system manufacturer [Referenz-ID: 583]
📐 Calibration from 9999-01-01
Blank substraction Applied
Zero gas generation Catalyst
Calibration frequency 1.0 months
Zero gas blank frequency 1.0 months
Linearity check frequency 12.0 months
Zero gas generation if catalyst
Catalyst temperature 400.0 degC

Calibration Certificates (1)

Name: NPL_NMHCs_2014_CMN Supplier: NPL Cylinder: D99 4206 Valid: 2014-03-11 → 2016-03-10
🧪 Air Treatment from 9999-01-01
Bool other
Co2 removal None
Water removal Nafion
Particle removal Stainless steel screen
📤 Data Submission
Databases
Database Submission
AGAGE 1.0 days
✅ Data QA/QC
Bool voc
Bool other
Bool ratios
Bool calibrations
📋 Measurement QA/QC
Checklist 1.0 weeks
Instrument log book Electronic
🧬 Compounds from 9999-01-01