NMHCs_GC_online_retired_CMN

retired
🔬 gas chromatograph, online 🏭 Agilent 📦 Agilent 6850
Description

Gas chromatograph operated at the observatory platform CMN-PV to perform online in-situ measurements of non-methane hydrocarbons (NMHCs).


🔧 Inlet System from 9999-01-01
Operated in atmospheric simulation chamber
Measurement height above ground 2.0 m
Measurement height above building 7.0 m
Distance from NOx inlet 0.0 m
Distance from ozone inlet 0.0 m
Inlet tube material composed inlet: 1 m of Pyrex (main inlet for all the instruments) then SS tube
Residence time 6.0 s
Inlet line heating
Inlet line temperature 20.0 °C
Filter inlet material Stainless steel
Lower mesh size 7.0 µm
💨 Sample from 9999-01-01
Sample air flow Vacuum sampling
Pump
Kind of pump Membrane
Pump manufacturer KNF
Pump model KNF N86
Air flow 40.0 mL/min
🧬 Compounds from 9999-01-01
🧪 Air Treatment during sampling from 9999-01-01
Water removal Nafion
CO₂ removal None
Particle removal Stainless steel screen
🌬️ Air Sampling from 9999-01-01
Air sampling technique Concentrating on adsorbents
Concentrating on adsorbents
Adsorption material MARKES U-T503F-2S
Adsorption temperature -35.0 °C
Desorption temperature 310.0 °C
Air sampling system manufacturer Markes
Air sampling system model Markes UNITY 2
Air sampling duration 25.0 min
Air sampling volume 1.0 L
💉 Injection from 9999-01-01
Injection technique Thermodesorption
Thermodesorption mode direct injection into the GC column at ambient temp.
Injection split Splitless
⚗️ GC System from 9999-01-01
Instrument manufacturer Agilent
Instrument model Agilent 6850
First column
Length 60.0 m
Material Agilent GasPro
Inner diameter 320.0 µm
📡 Detection from 9999-01-01
Detection method Mass spectrometry
Mass spectrometry system
Mass spectrometry manufacturer Agilent
Mass spectrometry model Agilent 5975C
Type Quadrupole
Tune frequency Monthly
📐 Calibration from 9999-01-01
Zero gas generation technique Catalyst
Zero gas generation with catalyst
Catalyst temperature 400.0 °C
Calibration frequency 1.0 months
Zero gas / Blank frequency 1.0 months
Blank substraction Applied
Linearity check frequency 12.0 months

Certificates (1)

Name: NPL_NMHCs_2014_CMN Supplier: NPL Cylinder: D99 4206 Valid: 2014-03-11 → 2016-03-10
📋 Measurement QA/QC
Instrument log book Electronic
Checklist 1.0 weeks
✅ Data QA/QC
Time series analysis of calibrations
Time series analysis of compound ratios
@VOC@
📤 Data Submission
Database Submission interval
AGAGE 1.0 days