NMHCs_GC_online_xx_ZEP

operational
🔬 gas chromatograph, online 🏭 Agilent 📦 Agilent 6890N
Description

Gas chromatograph operated at the observatory platform Zeppelin to perform online in-situ measurements of non-methane hydrocarbons (NMHCs).


🔧 Inlet System from 9999-01-01
Operated in atmospheric simulation chamber
Measurement height above ground 15.0 m
Measurement height above building 10.0 m
Distance from ozone inlet 30.0 m
Inlet tube material Stainless steel
Inlet line heating
Inlet line temperature 5.0 °C
Filter inlet material Stainless steel
Lower mesh size 60.0 µm
Upper mesh size 60.0 µm
💨 Sample from 9999-01-01
Sample air flow Vacuum sampling
Pump
Kind of pump Membrane
Pump manufacturer KNF
Pump model KNF UN05 Neoprene
Air flow 100.0 mL/min
🧬 Compounds from 9999-01-01
#Name
1 2-methylbutane
2 2-methyl-propane
3 benzene
4 ethene
5 n-butane
6 n-pentane
7 propane
8 toluene
🧪 Air Treatment during sampling from 9999-01-01
Water removal Nafion
CO₂ removal Other
CO₂ removal Using two cryotraps and flush out the CO2 during transfer from trap1 to trap2
Oxidant removal None
Particle removal Stainless steel screen
🌬️ Air Sampling from 9999-01-01
Air sampling technique Concentrating on adsorbents
Concentrating on adsorbents
Adsorption material Medusa
Adsorption temperature -170.0 °C
Desorption temperature 100.0 °C
Air sampling duration 20.0 min
Air sampling volume 2.0 L
💉 Injection from 9999-01-01
Injection technique Thermodesorption
Thermodesorption mode Cryofocus
Injection split Splitless
⚗️ GC System from 9999-01-01
Instrument manufacturer Agilent
Instrument model Agilent 6890N
Pre column
Length 0.5 m
Weight 150.0 mg
Material MS-4A
Lower mesh size 100.0 µm
Upper mesh size 110.0 µm
First column
Length 60.0 m
Material GC-GasPro
Inner diameter 320.0 µm
Second column
Length 25.0 m
Material CP-PoraBOUND Q
Inner diameter 320.0 µm
📡 Detection from 9999-01-01
Detection method Mass spectrometry
Mass spectrometry system
Mass spectrometry manufacturer Agilent
Type Quadrupole
Tune frequency Monthly
📐 Calibration from 9999-01-01
Zero gas generation technique N/A
Calibration frequency 1.0 weeks
Zero gas / Blank frequency 1.0 weeks
Blank substraction Applied
📋 Measurement QA/QC
Instrument log book Electronic
Checklist 1.0 days
✅ Data QA/QC
Time series analysis of calibrations
Time series analysis of compound ratios
@VOC@
📤 Data Submission
Database Submission interval
ACTRIS 1.0 years