NMHCs_GC_online_CN12191070_CVO

CV01L_Agilent_GC-FID_7890A_G3440A

operational
🔬 gas chromatograph, online 🏭 Agilent 📦 Agilent 7890A S/N CN12191070 📌 21.11157/3pgha-09v23
▶ Description

Gas chromatograph operated at the observatory platform Cape Verde Atmospheric Observatory to perform online in-situ measurements of non-methane hydrocarbons (NMHCs).


▶ 🔧 Inlet System from 9999-01-01
Operated in atmospheric simulation chamber ✗
Measurement height above ground 7.5 m
Measurement height above building 5.7 m
Distance from NOx inlet 0.0 m
Distance from ozone inlet 0.0 m
Inlet tube material 13 m of 40 mm diameter heated glass folllowed by 2 m 1/4" PTFE
Residence time 3.0 s
Inlet line heating ✓
Inlet line temperature 40.0 °C
Filter inlet material Stainless steel
Lower mesh size 2.0 µm
Upper mesh size 2.0 µm
▶ 💨 Sample from 9999-01-01
Sample air flow Vacuum sampling
Pump
Kind of pump Membrane
Pump manufacturer KNF
Air flow 28.0 mL/min
▶ 🧬 Compounds from 9999-01-01
#Name
1 2-methylbutane
2 2-methyl-propane
3 benzene
4 ethane
5 ethene
6 ethyne
7 isoprene
8 n-butane
9 n-pentane
10 propane
11 propene
12 toluene
▶ 🧪 Air Treatment during sampling from 9999-01-01
Water removal Cryotrap
CO₂ removal None
Particle removal Stainless steel screen
▶ 🌬️ Air Sampling from 9999-01-01
Air sampling technique Concentrating on adsorbents
Concentrating on adsorbents
Adsorption material Markes International Ozone Precursor Trap
Adsorption temperature -30.0 °C
Desorption temperature 300.0 °C
Air sampling system manufacturer Markes
Air sampling system model Markes UNITY 2
Sampling system serial number U21811 (Unity2) and CIA8 -10183 (CIA8)
Air sampling duration 40.0 min
Air sampling volume 1.1 L
▶ 💉 Injection from 9999-01-01
Injection technique Thermodesorption
Thermodesorption mode Cryofocus
Injection split Splitless
▶ ⚗️ GC System from 9999-01-01
Instrument manufacturer Agilent
Instrument model Agilent 7890A
Instrument serial number CN12191070
First column
Length 50.0 m
Material Alumina PLOT
Film thickness 10.0 µm
Inner diameter 530.0 µm
Second column
Length 60.0 m
Material StabilWax
Film thickness 1.5 µm
Inner diameter 530.0 µm
▶ 📡 Detection from 9999-01-01
Detection method FID
Detection system
Manufacturer Agilent
Model Agilent 7890A
Serial number CN12191070
▶ 📐 Calibration from 9999-01-01
Zero gas generation technique TOC-625 Prker hannafin filtration
Calibration frequency 1.0 months
Zero gas / Blank frequency 1.0 months
Blank substraction Applied
Linearity check frequency 4.0 hours

Certificates (1)

Name: NPL_NMHCs_2021_CVO Supplier: NPL Cylinder: D933520 Valid: 2021-11-04 → 2025-11-03
▶ 📋 Measurement QA/QC
Instrument log book Electronic
Checklist 1.0 weeks
▶ ✅ Data QA/QC
Time series analysis of calibrations ✓
Time series analysis of compound ratios ✓
@VOC@ ✗
▶ 📤 Data Submission
Database Submission interval
ACTRIS 1.0 years
– –
▶ 👤 Operators
OperatorLevel
0000-0003-4493-4086 expert