Logo

NMHCs_GC_online_xx_CVO

operational
🔬 gas chromatograph, online

📝 Notes

* Sampling system model: Unity2 and CIA8 (U-CIA 8-2S) * Oxidant removal: Heated line * Blank substraction: Applied for benzene only * Linearity check frequency: Annual, or when changes are made to system

💨 Sample from 9999-01-01
Pump
Kind of pump Membrane
Pump manufacturer [Referenz-ID: 576]
Air flow 28.0 mL/min
Sample air flow select Vacuum sampling
🔧 Inlet System from 9999-01-01
Line heating 40.0 degC
Residence time 3.0 s
Lower mesh size 2.0 µm
Upper mesh size 2.0 µm
Inlet tube material Other
Filter inlet material Stainless steel
Distance from nox inlet 0.0 m
Inlet line heating bool
Distance from ozone inlet 0.0 m
Inlet tube material if other 13 m of 40 mm diameter heated glass folllowed by 2 m 1/4" PTFE
Measurement height above ground 7.5 m
Measurement height above building 5.7 m
⚗️ GC System from 9999-01-01
First column
Length 50.0 m
Material Alumina PLOT
Film thickness 10.0 µm
Inner diameter 530.0 µm
Second column
Length 60.0 m
Material StabilWax
Film thickness 1.5 µm
Inner diameter 530.0 µm
Instrument model [Referenz-ID: 906]
Pre column select box
Instrument manufacturer [Referenz-ID: 547]
Instrument serial number CN12191070
Second column select box
📡 Detection from 9999-01-01
Select FID
System
Detection model [Referenz-ID: 906]
Detection manufacturer [Referenz-ID: 547]
Detection serial number CN12191070
🌬️ Air Sampling from 9999-01-01
Sampling volume 1.1 L
Sampling duration 40.0 min
Sampling system model [Referenz-ID: 901]
Air sampling technique Concentrating on adsorbents
Concentrating on adsorbents
Adsorption material Markes International Ozone Precursor Trap
Adsorption temperature -30.0 degC
Desorption temperature 300.0 degC
Sampling system manufacturer [Referenz-ID: 583]
Sampling system serial number U21811 (Unity2) and CIA8 -10183 (CIA8)
📐 Calibration from 9999-01-01
Blank substraction Applied
Zero gas generation Other
Calibration frequency 1.0 months
Zero gas blank frequency 1.0 months
Linearity check frequency 4.0 hours
Zero gas generation if other TOC-625 Prker hannafin filtration

Calibration Certificates (1)

Name: NPL_NMHCs_2021_CVO Supplier: NPL Cylinder: D933520 Valid: 2021-11-04 → 2025-11-03
🧪 Air Treatment from 9999-01-01
Bool other
Co2 removal None
Water removal Cryotrap
Particle removal Stainless steel screen
📤 Data Submission
Databases
Database Submission
ACTRIS 1.0 years
✅ Data QA/QC
Bool voc
Bool other
Bool ratios
Bool calibrations
📋 Measurement QA/QC
Checklist 1.0 weeks
Instrument log book Electronic
🧬 Compounds from 9999-01-01
#NameFormulaID
1 2-methylbutane 166
2 2-methyl-propane 164
3 benzene 225
4 ethane 252
5 ethene 255
6 ethyne 269
7 isoprene 288
8 n-butane 324
9 n-pentane 336
10 propane 361
11 propene 363
12 toluene 380