NMHCs_GC_offline_CN2350A157_IMT-NE
GC COVO
operational
🔬 gas chromatograph, offline
🏭 Agilent
📦 Agilent 8890
S/N CN2350A157
▶ Description
Gas chromatograph operated at the CiGas central facility Institut Mines Télécom Nord Europe to perform offline in-situ measurements of non-methane hydrocarbons (NMHCs)
▶ 🔧 Inlet System from 2024-05-02
| Operated in atmospheric simulation chamber | ✗ |
| Inlet tube material | sulfinert-treated stainless steel line |
| Inlet line heating | ✗ |
▶ 💨 Sample from 2024-05-02
| Canister sampling | Vacuum sampling |
| Pump | |
| Kind of pump | Membrane |
| Pump manufacturer | KNF |
| Pump model | KNF N87TT 9E |
| Air flow | 18.0 mL/min |
| Time | 45.0 minutes |
| Canister type | Electropolished stainless steel canisters |
| Air sampling volume | 0.81 litre |
| Sampling interval | 1.5 hours |
| Canister volume | 6.0 litre |
| Canister pressure pre-sampling | 0.0 bar |
| Canister pressure post-sampling | -0.13 bar |
| Number of replicates | 1.0 |
| Storage time | 21.0 days |
▶ 🧬 Compounds from 2024-05-02
| # | Name |
|---|---|
| 1 | ethane |
| 2 | propane |
| 3 | n-butane |
| 4 | 2-methylbutane |
| 5 | n-pentane |
| 6 | n-hexane |
| 7 | ethene |
| 8 | propene |
| 9 | ethyne |
| 10 | (E)-2-butene |
| 11 | 1-butene |
| 12 | (Z)-2-butene |
| 13 | 1,3-butadiene |
| 14 | (2Z)-2-pentene |
| 15 | (2E)-2-pentene |
| 16 | 1-pentene |
| 17 | 2-methylpentane |
| 18 | isoprene |
| 19 | benzene |
| 20 | n-heptane |
| 21 | 2,2,4-trimethylpentane |
| 22 | n-octane |
| 23 | toluene |
| 24 | ethylbenzene |
| 25 | m/p-xylenes |
| 26 | o-xylene |
| 27 | methacrolein |
| 28 | acetone |
| 29 | methyl vinyl ketone |
| 30 | methyl ethyl ketone |
| 31 | ethanol |
| 32 | 2-propanol |
| 33 | acetaldehyde |
▶ 🧪 Air Treatment during canister sampling from 2024-05-02
| Water removal | Ab-/Adsorbent type |
| Oxidant removal | Cartridges/filter material |
| Cartridges/filter material | KI/Cu |
| Particle removal | Stainless steel screen |
▶ 💉 Introduction device from 2024-05-02
| Technique | Concentrating on adsorbents |
| Concentrating on adsorbents | |
| Adsorption temperature | -30.0 °C |
| Desorption temperature | 280.0 °C |
| Inlet mode | Thermodesorption |
| Thermodesorption mode | Cold trap (peltier) |
| Air sampling system manufacturer | Markes |
| Air sampling system model | Markes UNITY-xr |
| Sampling system serial number | GB00U40028-22/12 |
| Sampling flow | 18.0 mL/min |
| Air sampling duration | 45.0 min |
| Injection volume | 0.81 L |
| Injection interval | 90.0 min |
| Injection split | Split |
| Injection pressure | -0.13 bar |
▶ ⚗️ GC System from 2024-05-02
| Instrument manufacturer | Agilent |
| Instrument model | Agilent 8890 |
| Instrument serial number | CN2350A157 |
| First column | |
| Length | 10.0 m |
| Material | GS-Oxyplot |
| Film thickness | 10.0 µm |
| Inner diameter | 530.0 µm |
| Second column | |
| Length | 50.0 m |
| Material | CP-Al2O3Na2SO4 |
| Film thickness | 10.0 µm |
| Inner diameter | 530.0 µm |
▶ 📡 Detection from 2024-05-02
| Detection method | FID |
| Detection system | |
▶ 📐 Calibration from 2024-05-02
| Zero gas generation technique | Catalyst |
| Zero gas generation with catalyst | |
| Catalyst type | Pt catalyst |
| Catalyst temperature | 420.0 °C |
| Number of repetitions | 5.0 |
| Blank substraction | Applied |
Certificates (1)
Name: NPL_NMHCs_2022_IMT-NE
Supplier: NPL
Cylinder: D982551
Valid: 2022-06-10 → 2023-06-09
▶ 📋 Measurement QA/QC
| Instrument log book | Electronic |
▶ ✅ Data QA/QC
| Time series analysis of calibrations | ✓ |
| Time series analysis of compound ratios | ✓ |
| @VOC@ | ✓ |
▶ 📤 Data Submission
| Database | Submission interval |
|---|---|
| – | – |