Logo

OVOCs_PTR_115_IMT-NE

PTRMS ionicon

operational
🔬 proton transfer mass spectrometer 🏭 Ionicon Analytik 📦 Ionicon Analytics PTR-Qi-TOF-MS S/N 115 ⚡ E/N 130.0 Td 🕑 Spectra averaging time/time resolution 10.0 s
Description

Proton transfer mass spectrometer operated at the CiGas central facility Institut Mines Télécom Nord Europe to perform in-situ measurements of oxygenated volatile organic compounds (OVOCs).


🔧 Inlet System from 2017-05-01
Operated in atmospheric simulation chamber
Inlet tube material Silcosteel
Inlet line heating
💨 Sample from 2017-05-01
Air flow 100.0 mL/min
🧬 Compounds from 2017-05-01
#Name
1 acetaldehyde
2 ethanol
3 isoprene
4 acetone
5 acetonitrile
6 methyl vinyl ketone
7 methyl ethyl ketone
8 benzene
9 toluene
10 m-xylene
11 1,2,4-trimethylbenzene
12 3-carene
13 dimethyl sulfide
14 perfluorotributylamine
15 1,2,4-trifluorobenzene
16 hexamethylcyclotrisiloxane
17 octamethylcyclotetrasiloxane
18 decamethylcyclopentasiloxane
19 1,2,4-trichlorobenzene
📐 Calibration from 2021-04-01
Calibration unit
Ecu system Standalone dilution system
Standalone dilution system
Kind of dilution system GCU
Dilution manufacturer Ionicon Analytik
Dilution model Ionicon Analytics GCU
Dilution serial number GCU-s-12-03(022)
Zero gas generation technique Catalyst
Zero gas generation with catalyst
Catalyst type Pt catalyst
Catalyst temperature 350.0 °C
Zero gas / Blank frequency 3.0 hours
Blank substraction Applied
Calibration frequency 7.0 days
Linearity check frequency 1.0 months

Certificates (1)

Name: NPL_OVOCs_2023_IMT-NE Supplier: NPL Cylinder: D913490 Valid: 2023-11-27 → 2024-11-24 Dilution: 0.000999 – 0.01960784
📋 Measurement QA/QC
Instrument log book Electronic
✅ Data QA/QC
Time series analysis of calibrations
Time series analysis of compound ratios
@VOC@