Logo

OVOCs_PTR_P4Q-288_Melpitz

installed
🔬 proton transfer mass spectrometer 🏭 TOFwerk 📦 TOFwerk Vocus PTR-S-TOF S/N P4Q-288

📝 Description
💨 Sample from 9999-01-01
🔧 Inlet System from 9999-01-01
Inlet line heating
📐 Calibration from 9999-01-01
Calibration unit
Ecu system Integrated in the PTR instrument
Integrated
Zero gas generation Catalyst
Zero gas generation if catalyst
Catalyst temperature 400.0 °C
Blank substraction Applied
Calibration frequency 1.0 months
Linearity check frequency 1.0 months
📤 Data Submission
Databases
Database Submission
ACTRIS 1.0 years
EMEP 1.0 years
✅ Data QA/QC
Time series analysis of calibrations
Time series analysis of compound ratios
@VOC@
Bool other
📋 Measurement QA/QC
Instrument log book Electronic
Checklist 1.0 days
🧬 Compounds from 9999-01-01
#Name
1 isoprene
2 benzene
3 toluene
4 m/p-xylenes
5 monoterpenes