NOx_CLD_187_Melpitz

installed
🔬 chemiluminescence detector 🏭 Teledyne 📦 Teledyne T200UP S/N 187

📝 Notes

* Converter: Teledyne Integrated in CLD * Zero Air manufacturer/model: homemade * NO and NO2 Calibration Frequency: now manualy (approx 1 / year) will later move to an automatic appraoch with more regular calibration * On-side calibration is not yet availabe. An homemade system is under developpement. Lab tests were successfull and a frist version will be tested on-site this summer. Until then, the calibration is performed at our lab, meaning that the instrument is transported from the station to the lab once approx once a year or after any kind of maintenance of the instrument.

🔧 Inlet System
Inlet tube material Glass coated steel
Filter inlet material Other
Inlet filter if other Quartz coated with PTFE
Distance from voc inlet 1.6 m
Inlet line heating bool
Humidity temperature control None
Measurement height above ground 3.7 m
📐 Calibration
No
Type Cylinder/Dilution
No model [Referenz-ID: 921]
Automation Automatic
Certificates
No manufacturer [Referenz-ID: 574]
No serial number 903916
No2
Type GPT
No2 model [Referenz-ID: 921]
Automation Automatic
Certificates
No2 manufacturer [Referenz-ID: 574]
No2 serial number 903916
No check
Zero air Zero Air generator
No2 check
Uncertainty 0.1 µmol/mol
Concentration 10.07 µmol/mol
📤 Data Submission
Databases
Database Submission
ACTRIS 1.0 years
EMEP 1.0 years
✅ Data QA/QC
Bool voc
Bool other
Bool ratios
Bool calibrations
📋 Measurement QA/QC
Instrument log book Hard copy