Logo

NMHCs_GC_online_650N8102405_PAS

FI01L_Perkin_Elmer_GC_FID_SAM

operational
🔬 gas chromatograph, online 🏭 PerkinElmer 📦 PerkinElmer Clarus 500 S/N 650N8102405 📌 http://hdl.handle.net/21.11157/bkv6b-bn959
Description

Gas chromatograph operated at the observatory platform Pallas Atmosphere-Ecosystem Supersite to perform online in-situ measurements of non-methane hydrocarbons (NMHCs).


🔧 Inlet System from 2012-01-02
Operated in atmospheric simulation chamber
Measurement height above ground 7.0 m
Measurement height above building 4.0 m
Distance from NOx inlet 0.0 m
Distance from condensables inlet 0.0 m
Distance from ozone inlet 0.0 m
Inlet tube material Stainless steel (OD/ID 60/56 mm, length 20 m), flow 2280-2940 L/min + Heated silcosteel (OD 1/4, ID 1/8 inch, length 3.7 m), flow 0.6 L/min + Silcosteel (OD 1/8, ID 1/16 inch, length 0.5 m), flow 0.020 L/min
Residence time 2.0 s
Inlet line heating
Inlet line temperature 30.0 °C
Filter inlet material None
💨 Sample from 2012-01-02
Sample air flow Vacuum sampling
Pump
Kind of pump Membrane
Pump manufacturer KNF
Pump model KNF N86 laboport
Air flow 20.0 mL/min
🧬 Compounds from 2012-01-01
#Name
1 ethane
2 propane
3 2-methyl-propane
4 n-butane
5 ethyne
6 2-methylbutane
7 n-pentane
8 isoprene
9 benzene
🧪 Air Treatment during sampling from 2012-01-01
Water removal Nafion
CO₂ removal None
Oxidant removal None
Particle removal Larger particles are following the flowing the flow in the main inlet line of the station
Other removal(s) during sampling Silco steel net inside the thermal desorption instrument
🌬️ Air Sampling from 2012-01-02
Air sampling technique Collection into cold trap
Air sampling system manufacturer PerkinElmer
Air sampling system model PerkinElmer Clarus 500
Sampling system serial number TD300L0810132
Sampling interval 12.0 hours
Air sampling duration 30.0 min
Air sampling volume 0.6 L
💉 Injection from 2012-01-01
Injection technique Thermodesorption
Thermodesorption mode Cryofocus
Cryofocus adsorption -40.0 °C
Cryofocus desorption 300.0 °C
Injection split Split
Percentage 80.0 %
⚗️ GC System from 2012-06-15
Instrument manufacturer PerkinElmer
Instrument model PerkinElmer Clarus 500
Instrument serial number 650N8102405
First column
Length 50.0 m
Material Al2O3/Na2SO4
Film thickness 5.0 µm
Inner diameter 320.0 µm
Second column
Length 60.0 m
Material BP5MS
Film thickness 1.0 µm
Inner diameter 250.0 µm
📡 Detection from 2012-01-02
Detection method FID
Detection system
Manufacturer PerkinElmer
Model PerkinElmer Clarus 500
Serial number 650N8102405
📐 Calibration from 2020-09-08
Zero gas generation technique Nitrogen 99.9999% gas in cylinder
Calibration frequency 3.0 weeks
Number of repetitions 5.0
Zero gas / Blank frequency 3.0 weeks
Blank substraction Applied
Linearity check frequency 12.0 months

Certificates (2)

Name: NPL_NMHCs_2020_PAS Supplier: NPL Cylinder: D723269 Valid: 2015-12-11 → 2017-12-10 Applied: 2020-09-09 → 2024-02-29
Name: NPL_NMHCs_2024_PAS Supplier: NPL Cylinder: D156916 Valid: 2024-01-18 → 2029-01-17 Applied: 2024-03-01 → 2029-01-18
📋 Measurement QA/QC
Instrument log book Electronic
Checklist 1.0 days
✅ Data QA/QC
Time series analysis of calibrations
Time series analysis of compound ratios
@VOC@
📤 Data Submission
Database Submission interval
ACTRIS
EMEP
GAW-WDCRG
👤 Operators
OperatorLevel
0000-0001-7022-3857 scientist