Logo

Condensables_CI-APi-TOF_xx_EUPHORE

operational
🔬 chemical ionization-atmospheric pressure interface-time of flight mass spectrometer
Description

Chemical ionization-atmospheric pressure interface-time of flight mass spectrometer operated at the atmospheric simulation chamber EUPHORE to perform in-situ measurements of condensable vapours.


🔧 Inlet System from 9999-01-01
Operated in atmospheric simulation chamber
Distance from NOx inlet 1.5 m
Distance from VOC inlet 1.5 m
Distance from ozone inlet 2.0 m
Inlet tube material PTFE
Residence time 1.7 s
Inlet line heating
Filter inlet material None
🧬 Compounds from 9999-01-01
🔬 Mass analyzer from 9999-01-01
Type Time-of-Flight
Time-of-Flight type L-TOF
Mass resolution 4000-6500
Instrument manufacturer Aerodyne Research
Reagent ions Iodide
Filter inlets for simultaneous measurements of gas and condensed phase HOM
Filter FIGAERO
Sample flow 2.0 l/min
Total air flow 3.3 l/min
Sample flow treatment with ion filter
📐 Calibration from 9999-01-01
Method Injections into chamber
Frequency When needed
📋 Measurement QA/QC
Instrument log book Hard copy
Checklist 1.0 days
✅ Data QA/QC
Time series analysis of calibrations
Time series analysis of compound ratios
@VOC@
📤 Data Submission
Database Submission interval
ACTRIS 1.0 years