NOx_CLD_02_EUPHORE

operational
🔬 chemiluminescence detector 🏭 Teledyne 📦 Teledyne T200UP
Description

Chemiluminescence detector operated at the atmospheric simulation chamber EUPHORE to perform in-situ measurements of nitrogen oxides (NOx).


🔧 Inlet System
Operated in atmospheric simulation chamber
Distance from VOC inlet 2.0 m
Distance from condensables inlet 0.0 m
Distance from ozone inlet 0.0 m
Inlet tube material PTFE (Teflon)
Residence time 20.0 s
Humidity/Temperature control Humidification with temperature condition at ambient
Inlet line heating
Filter inlet material PTFE (Teflon)
Inlet filter mesh size 5.0 µm
⇄ Converter
Type Photolytic
Converter manufacturer Teledyne
📐 Calibration
Gas standard concentration 20.0 µmol/mol
Zero air Cylinder
📏 Auxiliary Measurements
Role Type Manufacturer Model Serial number
Ozone instrument ultra-violet absorption detector Sabio Sabio 4010
Meteorology instrument Walz Walz Ts-2
📋 Measurement QA/QC
Instrument log book Hard copy
Checklist 1.0 days
✅ Data QA/QC
Time series analysis of calibrations
Time series analysis of compound ratios
@VOC@
📤 Data Submission
Database Submission interval
ACTRIS 1.0 years