Logo

NMHCs_GC_online_xx_EUPHORE

installed
🔬 gas chromatograph, online

📝 Description
💨 Sample
Sample air flow Vacuum sampling
Pump
Air flow 10.0 mL/min
🔧 Inlet System
Measurement height above ground 2.0 m
Distance from NOx inlet 3.0 m
Distance from ozone inlet 3.0 m
Inlet tube material Other
Inlet tube material stainless steel
Residence time 3.0 s
Inlet line heating
Line heating 80.0 °C
Filter inlet material None
⚗️ GC System
Instrument manufacturer Agilent
First column
Length 30.0 m
Material DB 5
Film thickness 250.0 µm
Inner diameter 250.0 µm
📡 Detection
Detection method Mass spectrometry
Mass spectrometry
Mass spectrometry manufacturer Agilent
🌬️ Air Sampling
Air sampling technique Other
Air sampling technique if other SPME
📐 Calibration
Zero gas generation Syn. Air supply
Zero gas blank frequency 1.0 days
🧪 Air Treatment
Water removal None
CO2 removal None
Oxidant removal None
Particle removal None
Bool other
📤 Data Submission
Databases
Database Submission
ACTRIS 1.0 years
✅ Data QA/QC
Time series analysis of calibrations
Time series analysis of compound ratios
@VOC@
Bool other
📋 Measurement QA/QC
Instrument log book Hard copy
Checklist 1.0 weeks
🧬 Compounds
#Name
1 ethane
2 propane
3 n-butane
4 n-pentane
5 n-hexane
6 n-heptane