Logo

OVOCs_PTR_Fusion_FZJV

Gas-Fusion

operational
🔬 proton transfer mass spectrometer 🏭 Ionicon Analytik 📦 Ionicon Analytik High-Sensitivity PTR-MS S/N FT10-023 ⚡ E/N 91.0 Td 🕑 Spectra averaging time/time resolution 1.0 s
Description

Proton-Transfer-Reaction-Time-of-Flight Mass Spectrometer with an online-GC option for the seperation of isomeric VOC


🔧 Inlet System from 2026-01-01
Operated in atmospheric simulation chamber
Measurement height above ground 2.0 m
Measurement height above building 1.0 m
Distance from NOx inlet 0.0 m
Distance from condensables inlet 0.0 m
Distance from ozone inlet 0.0 m
Distance from gc inlet 0.0 m
Inlet tube material e-polished stainless steel, heated > 70°C
Residence time 4.7 s
Inlet line heating
Inlet line temperature 90.0 °C
Filter inlet material None
Oxidant removal e-polished stainless steel, heated > 70°C
💨 Sample from 2026-04-01
Sample flow type Vacuum sampling
Sample flow rate 50.0 mL/min
Air flow 500.0 mL/min
Residence 2.0 s
🧬 Compounds from 2024-07-01
#Name
1 toluene
2 benzene
3 methyl vinyl ketone
4 acetone
5 m-xylene
6 3-carene
7 isoprene
8 decamethylcyclopentasiloxane
9 octamethylcyclotetrasiloxane
10 hexamethylcyclotrisiloxane
11 dimethyl sulfide
12 methyl ethyl ketone
📐 Calibration from 2025-03-31
Calibration unit
Ecu system Standalone dilution system
Dilution gas max flow 1.0 l/min
Diluent gas max flow 0.02 l/min
Standalone dilution system
Kind of dilution system LCU
Dilution manufacturer Ionicon Analytik
Zero gas generation technique Syn. Air supply
Zero gas / Blank frequency 1.0 hours
Blank substraction Applied
Calibration frequency 24.0 hours
Linearity check frequency 24.0 hours

Certificates (1)

Name: NPL_OVOCs_FZJV Supplier: NPL Cylinder: D296770 Valid: 2026-01-08 → 2027-01-07 Applied: 2025-03-01 → 2026-03-05 Dilution: 0.001 – 0.008
📋 Measurement QA/QC
Instrument log book Electronic
✅ Data QA/QC
Time series analysis of calibrations
Time series analysis of compound ratios
@VOC@
📤 Data Submission
Database Submission interval
ACTRIS