NMHCs_GC_online_Medusa_JFJ

operational
🔬 gas chromatograph, online

📝 Description
💨 Sample from 2008-01-01
Sample air flow Vacuum sampling
Pump
Kind of pump Membrane
Pump manufacturer KNF
Pump model KNF N86
Air flow 100.0 mL/min
🔧 Inlet System from 2008-01-01
Measurement height above ground 2.0 m
Measurement height above building -10.0 m
Distance from NOx inlet 50.0 m
Distance from ozone inlet 50.0 m
Inlet tube material Other
Inlet tube material not specified
Residence time 62.0 s
Inlet line heating
Line heating 10.0 °C
Filter inlet material Stainless steel
Lower mesh size 0.5 µm
Upper mesh size 0.5 µm
⚗️ GC System from 2008-01-01
Instrument manufacturer Agilent
Instrument model Agilent 6890
Instrument serial number CN10641022
First column
Length 60.0 m
Material GS-GASPRO
Film thickness 0.05 µm
Inner diameter 320.0 µm
Second column
Length 25.0 m
Material CP-PoraBOND Q
Film thickness 0.45 µm
Inner diameter 320.0 µm
📡 Detection from 2008-01-01
Detection method Mass spectrometry
Mass spectrometry
Mass spectrometry manufacturer Agilent
Mass spectrometry model Agilent 5975B
Mass spectrometry serial number US62734337
Type Quadrupole
Tune freq Monthly
🌬️ Air Sampling from 2008-01-01
Air sampling technique Concentrating on adsorbents
Concentrating on adsorbents
Adsorption material Hayesep
Adsorption temperature -160.0 °C
Desorption temperature 100.0 °C
Air sampling system manufacturer undefined
Air sampling system model undefined
Sampling system serial number Medusa-12
Air sampling duration 20.0 min
Air sampling volume 2000.0 L
📐 Calibration from 2008-01-01
Zero gas generation Other
Zero gas generation Alltec
Calibration frequency 2.0 hours
Number of repetitions 1.0
Zero gas blank frequency 1.0 weeks
Blank substraction Applied
Linearity check frequency 12.0 months

Calibration Certificates (1)

Name: NPL_NMHCs_2014_JFJ Supplier: NPL Cylinder: D99 4220 Valid: 2014-03-03 → 2026-03-02
🧪 Air Treatment from 2008-01-01
Water removal Nafion
CO2 removal None
Oxidant removal None
Particle removal Stainless steel screen
Bool other
📤 Data Submission
Databases
Database Submission
ACTRIS 1.0 years
✅ Data QA/QC
Time series analysis of calibrations
Time series analysis of compound ratios
@VOC@
Bool other
📋 Measurement QA/QC
Instrument log book Electronic
Checklist 1.0 weeks
🧬 Compounds from 2008-01-01
#Name
1 2-methylbutane
2 2-methyl-propane
3 benzene
4 ethane
5 propane
6 toluene
👤 Operators
OperatorLevel
User #21 scientist