NOx_CAPS_254_CESAM

installed
🔬 cavity attenuated phase shift spectrometer 🏭 Teledyne 📦 Teledyne T500U S/N 254
▶ Description

Cavity attenuated phase shift spectrometer operated at the atmospheric simulation chamber Multiphase Atmospheric Simulation Chamber (CESAM) to perform in-situ measurements of nitrogen oxides (NOx).


▶ 🔧 Inlet System
Operated in atmospheric simulation chamber ✓
Distance from VOC inlet 2.0 m
Distance from condensables inlet 1.0 m
Distance from ozone inlet 0.15 m
Inlet tube material PFA
Residence time 1.0 s
Humidity/Temperature control Manual control (see metadata)
Inlet line heating ✗
Filter inlet material Quartz
▶ 🎯 Zeroing
Zero air Cylinder
Frequency 1.0 days
Pressure 1000.0 mbar
▶ 📐 Calibration
Gas standard concentration 15.0 µmol/mol
Gas standard uncertainty 0.3 µmol/mol
Zero air Cylinder

NO Calibration

Type Cylinder/Dilution
Automation Manual
Frequency 12.0 months
NO calibrator manufacturer Thermo Fisher Scientific
NO calibrator model Thermo Fischer Scientific 146-i
NO calibrator serial number 1432863533

NO Certificates (1)

Name: Air_liquide_NOx_2016_CESAM Supplier: Other Applied: 2016-04-20 → 2016-09-01

NO₂ Calibration

Type GPT
Automation Manual
Frequency 12.0 months
NO₂ calibrator manufacturer Thermo Fisher Scientific
NO₂ calibrator model Thermo Fischer Scientific 146-i
NO₂ calibrator serial number 1432863533

NO₂ Certificates (1)

Name: Air_liquide_NOx_2016_CESAM Supplier: Other Applied: 2016-04-20 → 2026-09-01
▶ 📏 Auxiliary Measurements
Role Type Manufacturer Model Serial number
Ozone instrument ultra-violet absorption detector Horiba Horiba AP-370 9HVL0T8S
Meteorology instrument – Vaisala Vaisala HMP234 Y5010001
▶ 📋 Measurement QA/QC
Instrument log book Electronic
Checklist 1.0 days
▶ ✅ Data QA/QC
Time series analysis of calibrations ✓
Time series analysis of compound ratios ✗
@VOC@ ✗
▶ 📤 Data Submission
Database Submission interval
ACTRIS –