Logo

OVOCs_PTR_1664_CESAM

installed
🔬 proton transfer mass spectrometer 🏭 Kore 📦 Kore PTR-TOF 10k S/N 1664-1 ; 1664-2 ⚡ E/N 120.0 Td
▶ Description

Proton transfer mass spectrometer operated at the atmospheric simulation chamber Multiphase Atmospheric Simulation Chamber (CESAM) to perform in-situ measurements of oxygenated volatile organic compounds (OVOCs).


▶ 🔧 Inlet System from 9999-01-01
Operated in atmospheric simulation chamber ✓
Distance from NOx inlet 2.0 m
Distance from condensables inlet 1.0 m
Distance from ozone inlet 2.0 m
Inlet tube material silcosteel
Residence time 1.0 s
Inlet line heating ✓
Inlet line temperature 60.0 °C
▶ 💨 Sample from 9999-01-01
Sample flow type Vacuum sampling
Air flow 5.0 mL/min
Residence 1.0 s
▶ 🧬 Compounds from 9999-01-01
#Name
1 methanol
2 acetonitrile
3 acetaldehyde
4 ethanol
5 acetone
6 dimethyl sulfide
7 isoprene
8 methyl vinyl ketone
9 methyl ethyl ketone
10 benzene
11 toluene
12 m-xylene
13 1,2,4-trichlorobenzene
14 3-carene
15 1,2,4-trichlorobenzene
▶ 📐 Calibration from 9999-01-01
Calibration unit
Zero gas generation technique Syn. Air supply
Zero gas / Blank frequency 1.0 days
Blank substraction Applied
Calibration frequency 4.0 months
Linearity check frequency 4.0 months

Certificates (1)

Name: NPL_OVOCs_yearValidFrom_CESAM Supplier: NPL Applied: 2021-09-10 → 2026-09-10 Dilution: 10 – 20
▶ 📋 Measurement QA/QC
Instrument log book Electronic
Checklist 1.0 days
▶ ✅ Data QA/QC
Time series analysis of calibrations ✗
Time series analysis of compound ratios ✗
@VOC@ ✓
▶ 📤 Data Submission
Database Submission interval
ACTRIS –