Logo

NMHCs_GC_online_AgilentGC7890_FID _HPB

installed
🔬 gas chromatograph, online 🏭 Agilent 📦 Agilent 7890A 📌 21.11157/qyprc-wvk71
▶ Description

Gas chromatograph operated at the observational platform Meteorological Observatory Hohenpeissenberg (DWD) to perform online in-situ measurements of non-methane hydrocarbons (NMHCs).


▶ 🔧 Inlet System from 2026-01-01
Operated in atmospheric simulation chamber ✗
Measurement height above ground 17.0 m
Measurement height above building 2.0 m
Distance from NOx inlet 6.0 m
Distance from condensables inlet 5.0 m
Distance from ozone inlet 6.0 m
Inlet tube material Highflow glass manifold (ID ~4cm, ~12m/s) into the laboratory from there treated stainless steel 1/16" (~50ml/min)
Residence time 2.0 s
Inlet line heating ✗
Filter inlet material None
▶ 💨 Sample from 2026-01-01
Sample air flow Vacuum sampling
Pump
Kind of pump Membrane
▶ 🧬 Compounds from 9999-01-01
▶ 🧪 Air Treatment during sampling from 2026-01-01
Water removal Cryotrap
CO₂ removal None
Oxidant removal Cartridges/filter material
Cartridges/filter material Na2S2O3
Particle removal PTFE (Teflon)
▶ 🌬️ Air Sampling from 2026-01-01
Air sampling technique Concentrating on adsorbents
Concentrating on adsorbents
Adsorption material 1/4" treated stainless steel TD tube with CarbopackB and Carbosieve SIII
Adsorption temperature -25.0 °C
Desorption temperature 280.0 °C
Air sampling system manufacturer DWD
Air sampling system model DWD custom-made
Sampling system serial number 0003
Sampling interval 6.0 hours
Air sampling duration 15.0 min
Air sampling volume 0.8 L
▶ 💉 Injection from 2026-01-01
Injection technique Thermodesorption
Thermodesorption mode Direct desorption
Injection split Splitless
▶ ⚗️ GC System from 2026-01-01
Instrument manufacturer Agilent
Instrument model Agilent 7890A
First column
Length 50.0 m
Material AL2O3
▶ 📡 Detection from 2026-01-01
Detection method FID
Detection system
Manufacturer Agilent
▶ 📐 Calibration from 2026-01-01
Zero gas generation technique Catalyst
Zero gas generation with catalyst
Catalyst type Custom-made PD/PT catalyst
Catalyst temperature 400.0 °C
Calibration frequency 1.0 weeks
Number of repetitions 3.0
Zero gas / Blank frequency 1.0 days
Blank substraction Applied
Linearity check frequency 1.0 months

Certificates (2)

Name: NPL_NMHCs_2015_HPB Supplier: NPL Cylinder: D38 6624 Valid: 2015-12-09 → 2017-12-08 Applied: 2026-01-01 → ?
Name: NPL_NMHCs_2024_HPB Supplier: NPL Cylinder: D217056 Valid: 2024-07-05 → 2029-07-04 Applied: 2026-01-01 → ?
▶ 📋 Measurement QA/QC
Instrument log book Electronic
Checklist 1.0 days
▶ ✅ Data QA/QC
Time series analysis of calibrations ✓
Time series analysis of compound ratios ✓
@VOC@ ✓
Other QA/QC method time series of retention times and peak width
▶ 📤 Data Submission
Database Submission interval
ACTRIS 1.0 years
▶ 👤 Operators
OperatorLevel
0009-0006-5196-5402 scientist
0009-0009-1194-1926 scientist
Test User technician
Test User technician