NOx_CLD_xx_Vielsalm

operational
🔬 chemiluminescence detector 🏭 Horiba 📦 Horiba APNA-370CE S/N J7JFJ6P8
Description

Chemiluminescence detector operated at the observatory platform Vielsalm to perform in-situ measurements of nitrogen oxides (NOx).


🔧 Inlet System
Operated in atmospheric simulation chamber
Measurement height above ground 3.0 m
Measurement height above building 1.0 m
Distance from ozone inlet 0.2 m
Inlet tube material PTFE (Teflon)
Residence time 3.0 s
Humidity/Temperature control None
Inlet line heating
Inlet line temperature 20.0 °C
Filter inlet material PTFE (Teflon)
Inlet filter mesh size 5.0 µm
⇄ Converter
Type Molybdenum
Converter manufacturer Horiba
Converter model Horiba APNA-370CE
Converter serial number J7JFJ6P8
Duration of stay in converter or bypass line 0.5 s
Duration of stay in converter 0.5 s
Temperature 598.15 K
📐 Calibration
Gas standard concentration 700.0 µmol/mol
Gas standard uncertainty 14.0 µmol/mol
Gas standard confidence interval 95
Zero air Zero Air generator

NO Calibration

Type Cylinder/Dilution
Automation Manual
Frequency 3.0 months

NO₂ Calibration

Type GPT
Automation Manual
Frequency 3.0 months
NO₂ calibrator manufacturer Horiba
NO₂ calibrator model Horiba APMC-370
NO₂ calibrator serial number 903605
📏 Auxiliary Measurements
Role Type Manufacturer Model Serial number
Ozone instrument ultra-violet absorption detector Horiba Horiba AP-370 3B751ABX
Meteorology instrument
📋 Measurement QA/QC
Instrument log book Electronic
Checklist 1.0 days
✅ Data QA/QC
Time series analysis of calibrations
Time series analysis of compound ratios
@VOC@
📤 Data Submission
Database Submission interval
ACTRIS
EMEP 1.0 years