Logo

condensables_AI-API 006_ACD-C

operational
🔬 chemical ionization-atmospheric pressure interface-time of flight mass spectrometer
Description

Chemical ionization-atmospheric pressure interface-time of flight mass spectrometer operated at the atmospheric simulation chamber Atmospheric Chemistry Department Chamber (ACD-C) to perform in-situ measurements of condensable vapours.


🔧 Inlet System from 2024-01-01
Operated in atmospheric simulation chamber
Distance chamber wall 0.65 m
Distance from NOx inlet 1.9 m
Distance from VOC inlet 0.6 m
Distance from ozone inlet 1.7 m
Inlet tube material Glass
Residence time 1.4 s
Inlet line heating
Filter inlet material None
🧬 Compounds from 2024-01-01
#Name
1 sulfuric acid
🔬 Mass analyzer from 2024-01-01
Type Time-of-Flight
Time-of-Flight type C-TOF
Mass resolution 800-1200
Instrument manufacturer Aerodyne Research
Instrument model Aerodyne CI-API-TOF nitrate CIMS
Instrument serial number AI-API 006
Reagent ions Nitrate
Filter inlets for simultaneous measurements of gas and condensed phase HOM
Pressure 101300 Pa
Sample flow 12.0 l/min
Total air flow 12.0 l/min
Sample flow treatment with ion filter
📐 Calibration from 2024-01-01
Method Sulfuric acid calibrator
Frequency Once a year
📋 Measurement QA/QC
Instrument log book Electronic
Checklist 1.0 weeks
✅ Data QA/QC
Time series analysis of calibrations
Time series analysis of compound ratios
@VOC@
📤 Data Submission
Database Submission interval
ACTRIS 1.0 years
👤 Operators
OperatorLevel
0000-0002-3236-9780 scientist