NOx_CLD_12207616365_ACD-C

operational
🔬 chemiluminescence detector, molybdenum converter 🏭 Thermo Fisher Scientific 📦 Thermo 42i-QTL S/N 12207616365
▶ Description

Chemiluminescence detector with molybdenum converter operated at the observatory platform Atmospheric Chemistry Department Chamber (ACD-C) to perform in-situ measurements of nitrogen oxides (NOx).


▶ 🔧 Inlet System
Operated in atmospheric simulation chamber ✓
Distance chamber wall 0.65 m
Distance from VOC inlet 1.4 m
Distance from condensables inlet 1.9 m
Distance from ozone inlet 0.25 m
Inlet tube material PFA
Residence time 3.193114773 s
Humidity/Temperature control Manual control (see metadata)
Inlet line heating ✗
Filter inlet material PTFE (Teflon)
Inlet filter mesh size 0.5 µm
▶ ⇄ Converter
Type Molybdenum
Converter manufacturer Thermo Fisher Scientific
Converter model Thermo 42i-QTL
Converter serial number 12207616365
Temperature 598.0 K
▶ 📐 Calibration
Gas standard concentration 10.07 µmol/mol
Gas standard uncertainty 0.1 µmol/mol
Gas standard confidence interval 95%
Zero air Zero Air generator
Zero air manufacturer undefined
Zero air model undefined

NO Calibration

Type Cylinder/Dilution
Automation Automatic
NO calibrator manufacturer Horiba
NO calibrator model Horiba APMC-370
NO calibrator serial number 903916

NO₂ Calibration

Type GPT
Automation Automatic
NO₂ calibrator manufacturer Horiba
NO₂ calibrator model Horiba APMC-370
▶ 📏 Auxiliary Measurements
Role Type Manufacturer Model Serial number
Ozone instrument ultra-violet absorption detector Thermo Fisher Scientific Thermo 49iQ –
Meteorology instrument – Vaisala Vaisala HMT337 –
▶ 📋 Measurement QA/QC
Instrument log book Electronic
Checklist 1.0 weeks
▶ ✅ Data QA/QC
Time series analysis of calibrations ✓
Time series analysis of compound ratios ✗
@VOC@ ✗
▶ 📤 Data Submission
Database Submission interval
ACTRIS 1.0 years
▶ 👤 Operators
OperatorLevel
0000-0002-3236-9780 scientist