OVOCs_PTR_14-09-T8-048_ACD-C

operational
🔬 proton transfer mass spectrometer 🏭 Ionicon Analytik 📦 Ionicon Analytik PTR-TOF-8000 S/N 14-09-T8-048 🕑 Spectra averaging time/time resolution 10.0 s
Description

Proton transfer mass spectrometer operated at the atmospheric simulation chamber Atmospheric Chemistry Department Chamber (ACD-C) to perform in-situ measurements of oxygenated volatile organic compounds (OVOCs).


🔧 Inlet System from 2024-01-01
Operated in atmospheric simulation chamber
Distance chamber wall 0.65 m
Distance from NOx inlet 1.4 m
Distance from condensables inlet 0.6 m
Distance from ozone inlet 1.3 m
Inlet tube material PEEK
Residence time 0.628318531 s
Inlet line heating
Inlet line temperature 60.0 °C
Filter inlet material None
Oxidant removal None
💨 Sample from 2024-01-01
Sample flow type Vacuum sampling
Sample flow rate 300.0 mL/min
🧬 Compounds from 2026-06-01
#Name
1 isoprene
2 methyl vinyl ketone
3 benzene
4 alpha-pinene
5 acetonitrile
6 acetone
📐 Calibration from 2026-06-01
Calibration unit
Ecu system Custom-made dilution system
Dilution gas max flow 25.0 l/min
Diluent gas max flow 0.1 l/min
Custom
Diluent model undefined
Dilution model undefined
Diluent manufacturer undefined
Diluent serial number 02XX13 021839167
Dilution manufacturer undefined
Dilution serial number 06XX21 023125150
Zero gas generation technique In-house purified air supply
Zero gas / Blank frequency 1.0 days
Blank substraction Applied
Calibration frequency 1.0 months
Linearity check frequency 1.0 months

Certificates (1)

Name: Apfel-Riemer_OVOCs_ACD-C Supplier: Apel-Riemer Cylinder: D381974 Valid: 2026-06-09 → 2028-06-09 Applied: 2026-06-01 → ? Dilution: 4000 – 25
📋 Measurement QA/QC
Instrument log book Electronic
Checklist 1.0 weeks
✅ Data QA/QC
Time series analysis of calibrations
Time series analysis of compound ratios
@VOC@
📤 Data Submission
Database Submission interval
ACTRIS 1.0 years