Logo

NMHCs_GC_offline_CIAAdvantage_FZJV

operational
🔬 gas chromatograph, offline
Description

Gas chromatograph operated at the CiGas central facility Forschungszentrum Jülich - VOCs to perform offline in-situ measurements of non-methane hydrocarbons (NMHCs)


🔧 Inlet System from 9999-01-01
Operated in atmospheric simulation chamber
Measurement height above ground 5.0 m
Measurement height above building 2.0 m
Inlet tube material PTFE (Teflon)
Residence time 3.06 s
Inlet line heating
Filter inlet material PTFE (Teflon)
💨 Sample from 9999-01-01
Canister sampling Vacuum sampling
Pump
Kind of pump Metal below
Air flow 114.0 mL/min
Time 30.0 minutes
Air sampling volume 3.42 litre
Sampling interval 2.0 hours
Canister volume 6.0 litre
Canister pressure pre-sampling 0.0076 bar
Canister pressure post-sampling 0.537 bar
Number of replicates 1.0
Storage time 1.0 weeks
🧬 Compounds from 9999-01-01
#Name
1 ethane
2 ethene
3 ethyne
🧪 Air Treatment during canister sampling from 9999-01-01
Water removal None
CO₂ removal None
Particle removal PTFE (Teflon)
💉 Introduction device from 9999-01-02
Technique Concentrating on adsorbents
Concentrating on adsorbents
Inlet mode Thermodesorption
Thermodesorption mode Cold trap (peltier)
Air sampling system manufacturer Markes
Air sampling system model Markes UNITY-CIA Advantage-xr
Sampling system serial number GB00H25002-22/06, GB00W15006-22/06, GB00U40115-22/06
Sampling flow 50.0 mL/min
Air sampling duration 20.0 min
Injection volume 1.0 L
Injection interval 42.0 min
Injection split Split
Percentage 25.0 %
Injection pressure 1.137 bar
Number of replicates 3.0
⚗️ GC System from 9999-01-01
First column
Length 60.0 m
Material Dimethylpolysiloxane
Film thickness 1.0 µm
Inner diameter 320.0 µm
Second column
Length 50.0 m
Material Aluminium oxide
Film thickness 8.0 µm
Inner diameter 320.0 µm
📡 Detection from 9999-01-01
Detection method FID-MS
Detection system
Manufacturer Agilent
Model Agilent 8890
Serial number CN2233A063
Mass spectrometry system
📐 Calibration from 9999-01-01
Zero gas generation technique Syn. Air supply
Calibration frequency 1.0 weeks
Zero gas / Blank frequency 1.0 days
Linearity check frequency 1.0 weeks
📋 Measurement QA/QC
Instrument log book Electronic
Checklist 1.0 weeks
✅ Data QA/QC
Time series analysis of calibrations
Time series analysis of compound ratios
@VOC@
📤 Data Submission
Database Submission interval
ACTRIS
👤 Operators
OperatorLevel
0000-0002-6931-2622 technician
0000-0003-0648-6622 scientist