NMHCs_GC_offline_CIAAdvantage_FZJV

operational
🔬 gas chromatograph, offline

💨 Sample from 9999-01-01
Canister sampling Vacuum sampling
Pump
Kind of pump Metal below
Air flow 114.0 mL/min
Time 30.0 minutes
Air sampling volume 3.42 litre
Sampling interval 2.0 hours
Canister volume 6.0 litre
Canister pressure pre 0.0076 bar
Canister pressure post 0.537 bar
Number of replicates 1.0
Storage time 1.0 weeks
🔧 Inlet System from 9999-01-01
Measurement height above ground 5.0 m
Measurement height above building 2.0 m
Inlet tube material PTFE (Teflon)
Residence time 3.06 s
Inlet line heating
Filter inlet material PTFE (Teflon)
⚗️ GC System from 9999-01-01
First column
Length 60.0 m
Material Dimethylpolysiloxane
Film thickness 1.0 µm
Inner diameter 320.0 µm
Second column
Length 50.0 m
Material Aluminium oxide
Film thickness 8.0 µm
Inner diameter 320.0 µm
📡 Detection from 9999-01-01
Detection method FID-MS
System
Detection system manufacturer Agilent
Detection system model Agilent 8890
Detection system serial number CN2233A063
Mass spectrometry
📐 Calibration from 9999-01-01
Zero gas generation Syn. Air supply
Calibration frequency 1.0 weeks
Zero gas blank frequency 1.0 days
Linearity check frequency 1.0 weeks
🧪 Air Treatment from 9999-01-01
Water removal None
CO2 removal None
Particle removal PTFE (Teflon)
Bool other
📤 Data Submission
Databases
Database
ACTRIS
✅ Data QA/QC
Time series analysis of calibrations
Time series analysis of compound ratios
@VOC@
Bool other
📋 Measurement QA/QC
Instrument log book Electronic
Checklist 1.0 weeks
🧬 Compounds from 9999-01-01
#Name
1 ethane
2 ethene
3 ethyne
👤 Operators
OperatorLevel
User #5 technician
User #3 scientist